USER MANUAL
1. PRECAUTIONS AND SAFETY MEASUREMENTS
This instrument complies with safety Standard IEC/EN61010-1
related to electronic measuring instruments. For your own safety and
to avoid damaging the instrument follow the procedures described in
this instruction manual and read carefully all notes preceded by this
symbol
CAUTION
Avoid measuring in humid or wet places
Avoid measuring in rooms where explosive
gas, combustible gas, steam or excessive dust
is present
Keep you insulated from the object under test
Do not touch exposed metal parts such as test
lead ends, sockets, fixing objects, circuits etc.
Avoid doing that if you notice anomalous
conditions such as breakages, deformations,
fractures, leakages of battery fluid, blind display
etc.
Be careful when measuring on circuits with
voltages exceeding 20V to avoid risks of
electrical shock
The following symbols are used in user manual and on the meter:
CAUTION - refer to the instruction manual - an
improper use may damage the instrument or its
components
Double insulated meter
This symbol indicates that the clamp can operate on
live conductors
AC and DC current
Ground reference
CAUTION: this symbol indicates that equipment, its
accessories and battery shall be subject to a
separate collection and correct disposal
1.1 Preliminary instructions
CAUTION
This instrument has been designed for use in
environments with pollution degree 2
It can be used for AC and DC CURRENT
measurements on installations with
measurement category CAT III 300V
Do not test or connect to any circuit exceeding
the specified overload protection
Do not effect measurements under
environmental conditions exceeding the limits
indicated in § 11.2
Make sure that battery is properly installed
When the instrument is connected to
measuring circuits never touch any unused
terminal
If during measurement the displayed values
remain constant check whether the HOLD
function is active
1.2 Measuring overvoltage category definitions
According to the IEC/EN61010-1 guidelines the circuits are divided
into the following measurement categories:
CAT IV is for measurements performed at the source of the low-
voltage installation
Examples are electricity meters and measurements on primary
overcurrent protection devices and ripple control units
CAT III is for measurements performed in the building installation
Examples are measurements on distribution boards, circuit
breakers, wiring, including cables, bus-bars, junction boxes,
switches, socket-outlets in the fixed installation, and equipment
for industrial use and some other equipment, for example,
stationary motors with permanent connection to fixed installation
CAT II is for measurements performed on circuits directly
connected to the low voltage installation
Examples are measurements on household appliances, portable
tools and similar equipment
CAT I is for measurements performed on circuits not directly
connected to MAINS
Examples are measurements on circuits not derived from
MAINS, and specially protected (internal) MAINS-derived circuits.
In the latter case, transient stresses are variable; for that reason,
the standard requires that the transient withstand capability of the
equipment is made known to the user
2. GENERAL DESCRIPTION
The instrument has the following features:
DC/AC TRMS Current measurement up to 400A in Autorange
Inrush current measurement
Data HOLD and ZERO features
Backlight
Detection of AC Voltage without contact
Auto Power OFF
The selected quantity is displayed with indication of measuring
unit and active functions. The instrument is provided with an Auto
Power Off function consisting in an automatic switching off 15
minutes after last pressure on keys.
3. PREPARATION FOR USE
3.1 Initial checks
This instrument was checked both mechanically and electrically
prior to shipment. All possible cares and precautions were taken
to let you receive the instrument under perfect conditions.
Notwithstanding we suggest you to check it rapidly (any damage
may have occurred during transport – if so please contact the
local distributor from whom you bought the item).
Make sure that all standard accessories mentioned in § 11.2 are
included. Should you have to return back the instrument for any
reason please follow the instructions mentioned in § 12
3.2 Supply voltage
The instrument is powered by 2x1.5V battery type AAA IEC LR03
included on meter. When battery is low the symbol "
"is
displayed. To replace/insert the batteries follow the instructions
indicated in § 8
3.4 Storage
After a period of storage under extreme environmental conditions
exceeding the limits mentioned in § 11.2 let the instrument
resume normal operating conditions before using it
4. INSTRUMENT AND DISPLAY DESCRIPTION
5. SAFETY LINE
Put the cable as close as possible to the middle of the jaws
(considering the alignment marks) in order to meet the meter
accuracy specifications and take the hand under the safety line (see
below figure)
Copyright HT ITALIA 2017 Release EN 1.00 - 27/02/2017
6. FUNCTION KEYS DESCRIPTION
6.1
key
Press
key to switch on the instrument. Press and hold the key
to switch off the instrument
6.2 DH/
key
By pressing DH/
key the measured value is hold on the display
where the symbol "H" appear. Press again the key to disable this
function and resume normal operation. Press and hold the
DH/
key to activate/deactivate the display backlight
6.3 MODE key
The MODE key allows the selection of the type of current
measurement. The “AC” and “DC” symbols are shown at display
6.4 ZERO key
The ZERO key, which is active only in DC current measurement,
allows to activate/deactivate the zeroing of value at display in
order to remove the residual magnetization (see § 7.1) ). The “
”
symbol is shown at display
6.5 INRUSH key
The INRUSH key, which is active only in AC current measurement,
allows to activate/deactivate the Inrush current measurement
(instantaneous current peaks) of a installation. The “INRUSH”
symbol is shown at display and the 400A range is automatically
selected. Press and hold the INRUSH key to select the 40A range.
Press INRUSH key again to exit by the function
6.6 Detection of AC Voltage without contact
1. Switch on the instrument with
key
2. Move the NCV sensor (see figure) close to AC source
3. The switch on of red LED indicate the voltage presence
6.7 Auto Power OFF feature
In order to preserve internal batteries, the instrument switches
automatically off approximately 15 minutes after it was last used.
The “
” symbol appears on the display when this function is
active
7. HOW TO PERFORM THE MEASUREMENTS
7.1 DC and AC Current measurement
CAUTION
Put the conductor as close as possible to the
middle of the jaws in order to meet the meter
accuracy specifications
When apply/remove the instrument on/from live
conductors, if is not possible to de-energize the
circuit on test, use appropriate security
measures (e.g. protective gloves) to avoid
dangerous electric shock for the operator and a
possible damage of the instrument
Disconnect the main voltage or wear protective
gloves before performing measurement
Put the hand below the safety line during the
measurement as indicated in § 5
1. Switch on the instrument by pressing key
2. Press MODE key to select AC or DC measurement
3. For DC current measurement press the ZERO key in order to
remove the residual magnetization
4. Open the clamp jaw and insert the cable in the middle of it (see
figure – left part). Take care to the polarity for DC current
measurement
5. Do not perform measurements with more cables included in the
clamp jaw (see figure – right part)
6. The AC or DC current value will be shown on the display
7. If the message "OL" is displayed the maximum current value was
exceeded
8. For Data HOLD and INRUSH features see § 6.2 and § 6.5
8. BATTERY REPLACEMENT
When the "
" low battery indication is displayed the battery must
be replaced.
CAUTION
Only expert technicians should perform this operation.
Before carrying out this operation, make sure you
have removed the cable being tested from inside the
clamp jaw
1. Switch off the instrument by pressing
key
2. Remove the fixing screw and the battery compartment cover
3. Remove the batteries and replace them with new one of the same
type (see § 11.2) by respecting the correct polarity
4. Replace battery compartment cover
5. Do not scatter old batteries into the environment. Use the relevant
containers for disposal
9. CLEANING THE INSTRUMENT
Use a soft and dry cloth to clean the instrument. Never use wet cloths,
solvents, water, etc.
10. END OF LIFE
CAUTION: this symbol indicates that equipment, its
accessories and battery shall be subject to a
separate collection and correct disposal
11. TECHNICAL SPECIFICATIONS
11.1 Technical characteristics
Accuracy declared at temperature: 23°C 5°C, <80%RH
DC/AC TRMS Current
For not sinusoidal waveforms the frequency range is: 50Hz ÷60Hz
Overload protection: 400A AC
INRUSH current 40A range: 3A ÷ 40A, 50/60Hz, Accuracy: 10%rdg
INRUSH current 400A range:10A ÷ 400A, 50/60Hz, Accuracy: 10%rdg
INRUSH current response time: 100ms
11.2 General specifications
Reference guidelines
Safety: IEC/EN61010-1, EMC: IEC/EN61326-1
Measurement category: CAT III 300V
Pollution degree: 2, Insulation: double insulation
Max height of use: 2000m (6562ft)
Mechanical protection: IP30
Compliance with 2014/30/EU (EMC) and 2014/35/EU (LVD)
Display
Characteristics: LCD, 3 ¾dgt, 4000 points + decimal point
Conversion type: TRMS
Sampling rate: 3 times/s
Over range indication: “OL” symbol at display
Power supply
Battery type: 2x1.5V battery type AAA IEC LR03
Battery life: ca 85h (with backlight), ca 240h (without backlight)
Low battery indication: symbol “
” at display
Auto Power OFF: after 15 minutes of idleness
Environmental conditions for use
Working temp.: 540°C (41°F 104°F), <80%RH
Storage temperature: -20 60°C (-4°F 140°F), <80%RH
Mechanical characteristics
Dimensions (L x W x H): 155x60x25mm (6x2x1in)
Weight (with batteries): 140g (5 ounces)
Max diameter cable: 20mm (1in)
Standard accessories
Batteries
Carrying case
User manual
Range Resolution Frequency Accuracy
40.00A AC 0.01A
50Hz60Hz
(2.5%rdg+10dgt)
400.0A AC 0.1A
(2.8%rdg+8dgt)
40.00A AC 0.01A
60Hz400Hz
(2.8%rdg+10dgt)
400.0A AC 0.1A
(2.8%rdg+8dgt)
40.00A DC 0.01A
DC
(2.5%rdg+10dgt)
400.0A DC 0.1A
(2.8%rdg+8dgt)
12. SERVICE
12.1 Warranty conditions
This instrument is guaranteed for one year against material or
production defects, in accordance with our general sales conditions.
During the warranty period the manufacturer reserves the right to
decide either to repair or replace the product.
The warranty shall not apply in the following cases:
Repair and/or replacement of accessories and battery (not
covered by warranty).
Repairs that may become necessary as a consequence of an
incorrect use of the instrument or due to its use together with
non-compatible appliances.
Repairs that may become necessary as a consequence of
improper packaging.
Repairs which may become necessary as a consequence of
interventions performed by unauthorized personnel.
Modifications to the instrument performed without the
manufacturer’s explicit authorization
Use not provided for in the instrument’s specifications or in the
instruction manual
12.2 Service
Should the instrument not work properly, make sure that battery is
correctly installed and working and replace if necessary before
contacting your distributor. Should you need for any reason to return
back the instrument for repair or replacement take prior agreements
with the local distributor from whom you bought it. Do not forget to
enclose a report describing the reasons for returning (detected fault).
Use only original packaging. Any damage occurred in transit due to
not original packaging will be charged anyhow to the customer. The
manufacturer will not be responsible for any damage to persons or
things
HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – ITALY
Tel: +39-0546-621002
Fax: +39-0546-621144
Web: www.ht-instruments.com
ENG

Chatee con su manual de HT Instruments HT7005 usando IA

Obtenga respuestas instantáneas de IA de su manual de HT Instruments HT7005

Pregunte lo que quiera sobre HT Instruments HT7005. Las respuestas provienen de este manual, con las páginas en las que se basan.

HT7005 Programa de mantenimiento

Cuando se indique
Reemplazar bateria

Reemplace las baterias cuando aparezca el simbolo de bateria baja en la pantalla. Utilice dos pilas AAA de 1,5 V respetando la polaridad.

Mostrar en el manual (página 2)
Segun sea necesario
Limpiar instrumento

Utilice un pano seco suave para limpiar el instrumento. No utilice panos humedos, disolventes ni agua.

Mostrar en el manual (página 2)
M: QW-35-FL

Haz una Pregunta

0/1000

¿Recomiendas HT7005?

Sí No

Preguntas frecuentes

A

La HT7005 mide corrientes TRMS AC/DC hasta 400 A con auto-rango. Ejemplos de precisión: 40.00 A AC (50–60 Hz) ±(2.5% lectura + 10 dígitos); 400.0 A AC ±(2.8% lectura + 8 dígitos). La precisión en INRUSH es ±10% para los rangos INRUSH de 40 A y 400 A.

¿Esta respuesta fue útil?

A

Sí. La función INRUSH captura picos transitorios de arranque con rango automático de 400 A; mantenga pulsado para seleccionar el rango de 40 A. El tiempo de respuesta de INRUSH es aproximadamente 100 ms.

¿Esta respuesta fue útil?

A

Utiliza 2 × 1.5 V AAA (IEC LR03). La duración típica de las pilas es aproximadamente 85 horas con retroiluminación encendida y alrededor de 240 horas sin retroiluminación. Aparece un icono de batería baja en la pantalla cuando es necesario reemplazarlas.

¿Esta respuesta fue útil?

A

Sí, cuando se usa correctamente. El instrumento es de doble aislamiento, clasificado CAT III 300 V y conforme con IEC/EN61010-1. Siempre centre el conductor en la pinza, mantenga la mano bajo la referencia de seguridad y siga las pautas de EPP; evite mediciones en entornos húmedos, explosivos o con polvo excesivo.

¿Esta respuesta fue útil?

A

Rango de temperatura de funcionamiento 5–40 °C con <80% HR; almacenamiento −20 a 60 °C. Altitud máxima 2000 m. Protección mecánica IP30 y la pinza acepta hasta 20 mm de diámetro de cable.

¿Esta respuesta fue útil?

A

Use un paño suave y seco para la limpieza. No use agua, disolventes ni paños húmedos. Retire los conductores probados antes de reemplazar las pilas. Almacene el instrumento dentro de los límites ambientales especificados y déjelo volver a condiciones normales después de un almacenamiento extremo antes de usarlo.

¿Esta respuesta fue útil?

A

Los accesorios estándar incluyen pilas y una funda de transporte. Al final de su vida útil, el instrumento, los accesorios y las pilas deben desecharse por separado y gestionarse según las normativas locales de residuos electrónicos y de pilas.

¿Esta respuesta fue útil?

M: OA-GPT-5-NAN

Ventajas y Desventajas

Ventajas

  • Medición TRMS AC/DC precisa de corriente hasta 400 A con auto-rango.
  • El modo INRUSH captura transitorios de arranque con rangos seleccionables 40 A/400 A.
  • Detección de tensión sin contacto NCV y LCD retroiluminada para comprobaciones rápidas en campo.
  • Diseño compacto y ligero (140 g) con larga vida de pilas y apagado automático.
  • Clasificado según IEC/EN61010-1, doble aislamiento, CAT III 300 V, IP30.
VS

Desventajas

  • Apertura de la pinza limitada (20 mm) restringe el uso en conductores agrupados o grandes.
  • IP30 ofrece protección limitada contra polvo/agua—no apto para entornos húmedos exigentes.
  • La precisión INRUSH es ±10%, lo que puede ser insuficiente para análisis transitorios de alta precisión.
  • Acceso a las pilas recomendado solo para técnicos experimentados.
  • Sin registro avanzado ni interfaz para PC para análisis de datos extendido.
M: OA-GPT-5-NAN

Producto Especificación

Rango de medicion

max. 400A, max. 300V

Contenido del envio

Bolsa, Bateria

M: GO-GEM-FL-25

Producto Descripción

HT Instruments HT7005 Pinza Amperimétrica Multímetro Manual Resumen

La HT Instruments HT7005 es una pinza amperimétrica compacta y con auto-rango diseñada para mediciones TRMS precisas de corriente AC/DC hasta 400 A, con funciones adicionales orientadas al trabajo de campo y banco. Combina un diseño enfocado en la seguridad—cumpliendo IEC/EN61010-1 y clasificada CAT III 300 V—con funciones prácticas como captura de corriente de arranque (INRUSH), retención de datos, función ZERO para eliminar desplazamientos de CC y detección de tensión AC sin contacto. Su factor de forma ligero, pantalla LCD retroiluminada y larga vida de batería la hacen adecuada para electricistas, técnicos HVAC y personal de mantenimiento.

Características clave y uso

La HT7005 realiza lecturas TRMS de corriente tanto para formas de onda sinusoidales como no sinusoidales en sus rangos de frecuencia especificados, con capacidad de auto-rango en las escalas de 40 A y 400 A. La función INRUSH captura corrientes transitorias de arranque (rangos seleccionables 40 A/400 A) con un tiempo de respuesta de ~100 ms. Para mediciones en CC la función ZERO contrarresta la magnetización residual para mejorar la precisión en corrientes bajas. El sensor NCV (detección de tensión sin contacto) y el LED rojo permiten comprobaciones rápidas de presencia de tensión AC desde aproximadamente 200 V hasta 1000 V.

Se enfatiza la seguridad y el manejo adecuado: posicione el conductor en el centro de la pinza para la mejor precisión, mantenga las manos por debajo de la referencia de seguridad y evite mediciones en atmósferas húmedas o explosivas. El instrumento incluye protección contra sobrecarga hasta 400 A AC, doble aislamiento y protección mecánica IP30. Alimentado por dos pilas AAA, la HT7005 incluye apagado automático para conservar energía y muestra un icono de batería baja cuando es necesario reemplazarlas.

Consideraciones prácticas

La unidad se entrega con pilas y una funda de transporte, pesa aproximadamente 140 g con pilas y acepta conductores de hasta 20 mm de diámetro. Las condiciones típicas de funcionamiento son 5–40 °C, y está clasificada para uso hasta 2000 m de altitud. Las especificaciones de precisión varían según el rango y la frecuencia (por ejemplo, 40.00 A AC a 50–60 Hz: ±(2.5% lectura + 10 dígitos); 400.0 A AC: ±(2.8% lectura + 8 dígitos)). Siga los procedimientos de seguridad recomendados al probar conductores bajo tensión y evite insertar múltiples conductores dentro de la pinza.

Notas

  • Asegúrese de instalar las pilas con la polaridad correcta y reemplace ambas pilas AAA según sea necesario; solo personal cualificado debe abrir el compartimento de pilas.
  • Limpie con un paño suave y seco; evite disolventes y humedad.
  • Al final de su vida útil, deseche el dispositivo y las pilas conforme a las normativas locales de residuos electrónicos.